Benchtop pH / mV / ORP Meter, Five-Point Calibration

High-resolution benchtop pH and ORP meter reading to 0.001 pH with five-point calibration, selectable buffer sets and a calibration-due reminder, for laboratories working to a defined method.

Typical applications

  • Pharmaceutical and fine-chemical quality control
  • Research laboratories
  • ORP and redox measurement
  • Water and effluent analysis
  • Method development requiring a defined buffer set

Add to enquiry

Enquire on WhatsApp

Add as many products as you need and send one enquiry covering all of them, or send this item on its own. Bulk and drum quantities are quoted on request.

Worth knowing before you order

The ± 0.002 pH accuracy figure is the meter's, not the system's: it is only reachable with a sound electrode, calibrated over the range you actually measure, so the five calibration points and the calibration-due reminder are the useful part of this model rather than the extra decimal place. A general-purpose pH electrode, a temperature probe and buffers 4.01, 7.00 and 10.01 are supplied — but no ORP electrode is listed, and the ORP mode will not work without one, so order a redox electrode alongside if you need it. The 1.68 and 12.45 buffers of the five-point sets are also not supplied.

Specification

Technical specification for Benchtop pH / mV / ORP Meter, Five-Point Calibration
ModespH, mV, ORP, temperature
pH range−2.000 to 20.000 pH
pH accuracy± 0.002 pH
pH resolution0.001, 0.01, 0.1 pH, selectable
CalibrationUp to 5 points, with automatic buffer recognition
pH buffer setsUSA (1.68, 4.01, 7.00, 10.01, 12.45); NIST (1.68, 4.01, 6.86, 9.18, 12.45); DIN (1.09, 4.65, 6.79, 9.23, 12.7); or user defined
Calibration due reminder1 to 30 days or off
mV range± 1999 mV
mV accuracy± 0.2 mV
mV resolution0.1 or 1 mV, selectable
ORP modeAbsolute or relative millivolt, single-point offset calibration
Temperature range0–105 °C, 32–212 °F
Temperature accuracy± 0.5 °C, ± 0.9 °F
Temperature compensation0–100 °C, 32–212 °F, automatic or manual
Auto-offManual, or automatic 10 / 20 / 30 minutes after the last key press
Memory500 readings with date and time
Real-time clockBuilt in, date and time stamped on stored data
Power supplyDC adaptor from AC 220 V, 50/60 Hz
Packing dimensions (L x W x H)310 x 250 x 190 mm
Gross weight2 kg

Support available

  • Installation & commissioning
  • Calibration
  • Validation & qualification (IQ/OQ/PQ)
  • Operator training
  • Annual maintenance contract
  • Spares & consumables

Calibration is arranged through NABL-accredited laboratories under the relevant scope. We state the scope and the certificate number on the quotation — see what accreditation actually covers.