Indent

Probe Sonicator with Acoustic Enclosure, 650 to 1800 W

Four-model range of ultrasonic probe processors supplied with a lit, UV-equipped acoustic enclosure and a circulating beaker, covering working volumes from 0.2 mL to 1500 mL.

Typical applications

  • Disruption of animal, plant and virus cells
  • Emulsification and separation
  • Homogenisation and extraction
  • Nanoparticle dispersion
  • Defoaming and cleaning
  • Accelerating chemical reactions

Add to enquiry

Enquire on WhatsApp

Add as many products as you need and send one enquiry covering all of them, or send this item on its own. Bulk and drum quantities are quoted on request.

Worth knowing before you order

The probe decides what the machine can do; the generator only sets the ceiling. Each model ships with one horn — 6, 10, 20 and 22 mm as the power rises — and the probe table ties every diameter to a volume band and to a usable amplitude range, so an 1800 W unit on its 22 mm horn will not process 5 mL and a 650 W unit on the 2 mm microtip will not process 500 mL. Buy the probes for the volumes you actually run rather than the wattage that sounds safest. Sonication is a heating process: the energy ends up as heat in a small volume, so an ice bath and pulsed operation are normal practice, and the jacketed circulating beaker supplied as standard only circulates if you also buy a recirculating chiller to feed it, which is not part of the package. The enclosure is not a luxury — an unhoused horn at these powers is loud enough to damage hearing — and its UV lamp treats only the exposed surfaces of an empty cabinet. One inconsistency to settle on the quotation: the PRO656 minimum volume is printed as 0.2 mL in the specification table and 0.1 mL in the probe table.

Specification

Technical specification for Probe Sonicator with Acoustic Enclosure, 650 to 1800 W
ModelsPRO656, PRO900, PRO1200, PRO1800
Frequency, PRO656 / PRO90020 to 25 kHz
Frequency, PRO1200 / PRO180019.5 to 20.5 kHz
Power (adjustable), by modelPRO656 6.5 to 650 W, PRO900 9 to 900 W, PRO1200 12 to 1200 W, PRO1800 18 to 1800 W
Crushing capacity, by modelPRO656 0.2 to 500 mL, PRO900 0.5 to 700 mL, PRO1200 3 to 1000 mL, PRO1800 10 to 1500 mL
Standard probe, by modelPRO656 6 mm, PRO900 10 mm, PRO1200 20 mm, PRO1800 22 mm
Optional probes, by modelPRO656 2/3/6/8/10/12 mm, PRO900 6/8/10/12/15 mm, PRO1200 10/15/20/22 mm, PRO1800 6/8/10/15/20/25 mm
Volume by probe diameter, PRO6562 mm 0.1-5 mL, 3 mm 3-10 mL, 6 mm 10-100 mL, 8 mm 20-200 mL, 10 mm 30-300 mL, 12 mm 50-500 mL
Volume by probe diameter, PRO900As PRO656 from 2 to 12 mm, plus 15 mm 100-700 mL
Volume by probe diameter, PRO12003 mm 3-10 mL, 6 mm 10-100 mL, 8 mm 20-200 mL, 10 mm 30-300 mL, 15 mm 100-500 mL, 20 mm 100-1000 mL, 22 mm 200-1000 mL
Volume by probe diameter, PRO18006 mm 10-100 mL, 8 mm 20-200 mL, 10 mm 30-300 mL, 15 mm 100-500 mL, 20 mm 100-1000 mL, 22 mm 200-1500 mL, 25 mm 500-1500 mL
Usable amplitude band by probeEach probe has its own power ratio range, from 1-35 % on the 2 mm probe to 20-95 % on the 25 mm probe
Temperature range0 to 100 °C
Single ultrasound time and interval time0.1 to 99.9 s
Display7 inch colour touch screen, with pulse, frequency and time shown as a waveform
Data storage20 sets
Alarm systemTime, overload, no-load and over-temperature
Acoustic enclosureSoundproof enclosure with built-in LED light and ultraviolet sterilisation; 355 x 355 x 545 mm
Sample handlingCirculating glass beaker supplied as standard
Power supplyAC 220 V, 50/60 Hz
Packing dimensionsPRO656 490 x 320 x 450 mm, PRO900 480 x 320 x 450 mm, PRO1200 and PRO1800 490 x 325 x 450 mm
Gross weight, instrument and enclosure32 kg (22 kg + 10 kg)
HSN code84798200
Supplied withMain instrument, transducer with standard probe, temperature probe, power cable, LED lamp cable, soundproof enclosure, circulating glass beaker, instruction manual, test certificate

Support available

  • Installation & commissioning
  • Operator training
  • Annual maintenance contract
  • Spares & consumables

Calibration is arranged through NABL-accredited laboratories under the relevant scope. We state the scope and the certificate number on the quotation — see what accreditation actually covers.