Indent

Moisture Analyser, 160 g x 1 mg, Halogen, 5 in Touch Screen

Halogen loss-on-drying moisture analyser with a 5 in colour touch screen, multi-user password access and stored product methods, for determining moisture as the mass lost on heating.

Typical applications

  • Moisture in food, feed and agricultural products
  • Loss on drying of powders and granules
  • Moisture in plastics, resins and rubber compounds
  • Solids content of pastes, slurries and emulsions
  • Multi-product QC where each material has its own stored method

Add to enquiry

Enquire on WhatsApp

Add as many products as you need and send one enquiry covering all of them, or send this item on its own. Bulk and drum quantities are quoted on request.

Worth knowing before you order

The result is a loss-on-drying figure, so it belongs to the method: change the drying temperature, the profile or the end-point rule and the number changes with it. Fix and document one method per product before you start comparing batches, and do not treat the reading as interchangeable with an oven result until the two have been correlated. One conflict in the source is worth resolving before ordering: the feature list names four profiles (standard, fast, mild, steps) while the specification table lists only Standard and Rapid - if a ramped or stepped drying profile is part of your method, get that confirmed in writing. The ceiling is 160 °C, so hotter methods need a different model. Calibration is external and no calibration weight is listed in the accessories. Records stay on the instrument and come off over RS232C or USB; the printer or PC is a separate item.

Specification

Technical specification for Moisture Analyser, 160 g x 1 mg, Halogen, 5 in Touch Screen
Heating sourceHalogen lamp
Temperature setting40 °C to 160 °C, 1 °C step
Temperature profiles (specification table)Standard and Rapid
Weighing capacity160 g
Readability0.001 g (1 mg)
Repeatability0.001 g (1 mg)
Linearity (±)0.002 g
Moisture result resolution0.01 %
Results displayedMoisture %, dry residue %, ATRO, weight; live drying graph
Drying time setting1 to 99 min, or auto-stop
Display5 in colour touch screen, 800 x 480
Methods, records and access30-product database with 10 rapid keys; 3000 test reports and drying graphs per user; multi-user password protection
Printout fieldsDate, time, instrument ID, user ID, project ID, temperature, mode, moisture %, signature line
Sample pan size and weighing sensor100 mm; loadcell
CalibrationExternal; GLP/ISO-format record of the last stored calibration (manufacturer's description)
Operating temperature5 °C to 35 °C
InterfaceRS232C, USB
Power supplyAC 220 V / 50-60 Hz
Packing dimensions (L x W x H)450 x 285 x 390 mm
Gross weight7 kg
Supplied withAluminium pans (50 nos.), pan supporter, power cable, instruction manual, test certificate
HSN code90160010

Support available

  • Installation & commissioning
  • Calibration
  • Validation & qualification (IQ/OQ/PQ)
  • Operator training
  • Annual maintenance contract
  • Spares & consumables

Calibration is arranged through NABL-accredited laboratories under the relevant scope. We state the scope and the certificate number on the quotation — see what accreditation actually covers.