Visible Spectrophotometer, Automatic Wavelength Setting
Single-beam visible spectrophotometer covering 325 to 1000 nm at 4 nm bandwidth, with motorised wavelength setting, for colorimetric determinations against a standard curve.
Typical applications
- Colorimetric determinations against a standard curve
- Water and effluent testing
- Food and beverage quality control
- Clinical and biochemical colorimetry
- Teaching laboratories
Add to enquiry
Add as many products as you need and send one enquiry covering all of them, or send this item on its own. Bulk and drum quantities are quoted on request.
Worth knowing before you order
Visible only. There is no deuterium lamp and the range starts at 325 nm, so any UV assay or identification test below that wavelength — which is most of the pharmacopoeial UV work — cannot be run on this instrument. The 4 nm spectral bandwidth is wider than any narrow-band requirement a method is likely to state, and stray light is quoted at ≤ 0.3 %T, so treat absorbances above about 2 A with caution. It is easy to confuse with the manual-wavelength visible model, which looks much the same and carries the same 4 nm bandwidth and 325 to 1000 nm range: this one sets wavelength automatically and holds ± 1 nm accuracy and ± 0.5 nm repeatability, against a hand-set knob and ± 2 nm accuracy on the other. Confirm which one is on the order.Specification
| Beam configuration | Single beam, 1200 lines/mm grating |
|---|---|
| Wavelength range | 325–1000 nm |
| Spectral bandwidth | 4 nm |
| Wavelength accuracy | ± 1 nm |
| Wavelength repeatability | ± 0.5 nm |
| Wavelength setting | Automatic |
| Photometric accuracy | ± 0.5 %T |
| Photometric repeatability | ± 0.3 %T |
| Photometric range | −0.3 to 3 A; 0–200 %T |
| Photometric modes | T, A, C, F |
| Stray light | ≤ 0.3 %T |
| Stability | ± 0.002 A/h at 500 nm |
| Detector and light source | Silicon photodiode and tungsten lamp |
| Cell holder | Four-position 10 mm cell holder |
| Display and memory | 2.5 in LCD, five results per screen; stores up to 200 results |
| Data output | USB port and parallel printer port |
| Power supply | AC 220 V, 50/60 Hz |
| Packing dimensions (L x W x H) | 565 x 435 x 305 mm |
| Gross weight | 11 kg |
Support available
- Installation & commissioning
- Calibration
- Operator training
- Annual maintenance contract
- Spares & consumables
Calibration is arranged through NABL-accredited laboratories under the relevant scope. We state the scope and the certificate number on the quotation — see what accreditation actually covers.
Related products
Microvolume Spectrophotometer (Biophotometer)
Cuvette-free spectrophotometer that reads nucleic acid, protein, colorimetric and OD600 absorbance from a 0.5 to 2 µL drop placed directly on the sample pedestal.
IndentWavelength range: 190–850 nm
Specification for Microvolume Spectrophotometer (Biophotometer)
Specification for Microvolume Spectrophotometer (Biophotometer) Wavelength range 190–850 nm Sample volume 0.5–2 µL Optical path lengths, high concentration 0.02 mm, 0.05 mm Optical path lengths, general concentration 0.2 mm, 1.0 mm Nucleic acid testing range 2–27500 ng/µL (dsDNA) Absorbance range 0.02–3.00 A Absorbance accuracy 0.002 Abs at the 0.2 mm path Light source Xenon flash lamp Detector One UV-enhanced CMOS line array sensor OD600 absorbance range 0 to 6.000 Abs OD600 stability and repeatability (0.3) ≤ 0.5%, (3.4) ≤ 2% Fluorescence sensitivity dsDNA 0.5 pg/µL 9 further specifications on the full record.
Typical applications
- DNA and RNA quantification
- Protein quantification
- OD600 cell density of bacterial cultures
- Colorimetric assays
- Fluorescence quantitation of dsDNA and RNA
- Enzyme kinetics
Also known as
- Microvolume UV-Vis spectrophotometer
- Biophotometer
- Nucleic acid quantification spectrophotometer
- DNA and protein quantitation spectrophotometer
Full record, hazard data and enquiry for Microvolume Spectrophotometer (Biophotometer) →
PMT Double-Beam UV-Visible Spectrophotometer with Mercury Lamp
Double-beam UV-visible spectrophotometer with a photomultiplier detector and a spectral bandwidth continuously adjustable from 0.1 to 5 nm, controlled from PC software.
IndentUSP <857>Beam configuration: Double beam, Czerny-Turner optics
Specification for PMT Double-Beam UV-Visible Spectrophotometer with Mercury Lamp
Specification for PMT Double-Beam UV-Visible Spectrophotometer with Mercury Lamp Beam configuration Double beam, Czerny-Turner optics Detector Photomultiplier (PMT) Wavelength range 190–900 nm Spectral bandwidth 0.1–5 nm, continuously adjustable in 0.1 nm steps Resolution 0.1 nm Wavelength accuracy ± 0.3 nm Photometric accuracy ± 0.3% Photometric range, absorbance and concentration −4 A to 6 A; 0 to 9999 C Stray light ≤ 0.005% at 220 nm and 360 nm Stability ≤ 0.0001 Abs/h Light sources Deuterium, tungsten and mercury lamps Mercury lamp function Real-time wavelength calibration 8 further specifications on the full record.
Standards
USP <857>Typical applications
- Pharmaceutical assay and identification
- Samples with sharp absorption peaks, such as the penicillins
- Wavelength scanning
- Kinetics and time-course measurement
- Trace-level absorbance work
Also known as
- PMT spectrophotometer
- Double beam UV-Vis spectrophotometer with photomultiplier
- Variable bandwidth UV-Vis spectrophotometer
Touchscreen Double-Beam UV-Visible Spectrophotometer
Double-beam UV-visible spectrophotometer with a 10-inch touchscreen and on-board data handling, offered with fixed or selectable spectral bandwidth and with a single or eight-position cell holder.
IndentUSP <857>Beam configuration: Double beam
Specification for Touchscreen Double-Beam UV-Visible Spectrophotometer
Specification for Touchscreen Double-Beam UV-Visible Spectrophotometer Beam configuration Double beam Wavelength range 190–1100 nm Spectral bandwidth, first and second models 1.0 nm Spectral bandwidth, third and fourth models 1.0, 2.0, 4.0, 5.0 nm, selectable Sample cell holder, first and third models 10 mm cell holder Sample cell holder, second and fourth models 8-cell programmable holder, 10 mm Cell holder 10 mm each, sample and reference cuvette holder Wavelength accuracy and repeatability ± 0.1 nm at the D2 656.1 nm line, ± 0.3 nm full range; repeatability ± 0.1 nm Resolution 0.1 nm Photometric range 0–200 %T; −0.3 to 3.0 A; 0–9999 C Photometric accuracy ± 0.2 %T (0–100 %T); ± 0.002 Abs (0–0.5 Abs); ± 0.004 Abs (0.5–1.0 Abs) Stability ± 0.0004 A/h at 500 nm 9 further specifications on the full record.
Standards
USP <857>Typical applications
- Pharmaceutical assay and content uniformity
- Spectral scanning
- Kinetics and time-course measurement
- DNA and protein quantification
- Multi-wavelength analysis
- Quantitative analysis against calibration curves
Also known as
- Double beam UV-Vis spectrophotometer
- Touchscreen spectrophotometer
- Stand-alone double beam spectrophotometer
Full record, hazard data and enquiry for Touchscreen Double-Beam UV-Visible Spectrophotometer →
Touchscreen Single-Beam UV-Visible Spectrophotometer
Single-beam UV-visible spectrophotometer with a 7-inch touchscreen, 2 nm spectral bandwidth and on-board data handling, for fixed-wavelength assay and routine scanning.
IndentUSP <857>Beam configuration: Single beam
Specification for Touchscreen Single-Beam UV-Visible Spectrophotometer
Specification for Touchscreen Single-Beam UV-Visible Spectrophotometer Beam configuration Single beam Wavelength range 190–1100 nm Spectral bandwidth 2.0 nm, fixed Resolution 0.1 nm Wavelength accuracy ± 0.5 nm Wavelength repeatability ± 0.2 nm Photometric range 0–200 %T; −0.3 to 3.0 A; 0–9999 C Photometric accuracy ± 0.2 %T (0–100 %T); ± 0.002 Abs (0–0.5 Abs); ± 0.004 Abs (0.5–1.0 Abs) Photometric repeatability ≤ 0.1 %T (0–100 %T); 0.001 Abs (0–0.5 Abs); 0.002 Abs (0.5–1.0 Abs) Stability ± 0.001 A/h at 500 nm Stray light 0.05 %T at 220 nm and 360 nm Cell holder 10 mm each, sample and reference cuvette holder 9 further specifications on the full record.
Standards
USP <857>Typical applications
- Fixed-wavelength quantitative assay
- Colorimetric determinations in water and effluent
- Spectral scanning
- DNA and protein quantification
- Multi-wavelength analysis
- Teaching and routine quality control
Also known as
- Single beam UV-Vis spectrophotometer
- Touchscreen single beam spectrophotometer
- Stand-alone UV-Vis spectrophotometer
Full record, hazard data and enquiry for Touchscreen Single-Beam UV-Visible Spectrophotometer →