Glass & Glass Containers

Testing for flat, container and pharmaceutical glass — thickness and stress, fragmentation, thermal shock and hydrolytic resistance.

"Type I glass" is not a material specification. It is the outcome of a titration. Hydrolytic resistance is measured by extracting crushed glass in water under defined conditions — ISO 719 at 98 °C, ISO 720 at 121 °C in an autoclave — and titrating the extract against dilute acid, and the class follows from the titre. Which makes the test only as good as the water: the extraction water has to meet the purity the method specifies, and ordinary distilled water standing in a soda-glass carboy will leach enough alkali to move the result on its own. The flat-glass side of the sector has the opposite problem — its acceptance test, fragmentation counting, destroys the pane, so the useful daily measurement is non-destructive surface stress taken before the toughening furnace, where something can still be done about it.

Tests run here

  • Hydrolytic resistance of glass grains and of container inner surfaces
  • Surface and edge stress by photoelastic measurement
  • Fragmentation count on thermally toughened glass
  • Thermal shock and thermal endurance
  • Thickness, bow and optical distortion on flat glass
  • Internal pressure and vertical load resistance of containers
  • Residual strain by polariscope
  • Light transmission and arsenic on pharmaceutical containers

Governed by

  • IS 2553 (Part 1) — Safety glass for architectural, building and general uses
  • IS 14900 — Transparent float glass, specification
  • ISO 719 and ISO 720 — hydrolytic resistance of glass grains
  • ASTM C1036 for flat glass and ASTM C1048 for heat-treated flat glass
  • EN 12150 — thermally toughened soda lime silicate safety glass
  • Indian Pharmacopoeia and USP requirements for glass containers
  • NABL accreditation under ISO/IEC 17025